Side-by-side comparison
| Upfront installation | Ducted: hood + blower + duct + roof discharge. Ductless: hood only, no ductwork. |
|---|---|
| Chemical range | Ducted: broad, subject to the exhaust design. Ductless: limited to chemicals the filter is rated for. |
| Recurring cost | Ducted: fan energy and maintenance. Ductless: filter replacement at intervals set by usage. |
| Heat and solvent load | Ducted handles high loads. Ductless returns air and heat to the room. |
| Compliance | Both need their performance verified. Ductless additionally needs a documented filter management process. |
| Relocation | Ducted: tied to the building exhaust. Ductless: easier to move. |
When ductless is the right answer
A teaching or demonstration room with a short, stable chemical list; a rented building where ductwork is not permitted; or a temporary installation. In all three cases, write the permitted chemical list into the operating procedure and keep a filter log.
When ducted is the right answer
Research and industrial labs where procedures change, where solvents or high heat are used, or where the project standard requires discharge to an external exhaust. This covers most new-build laboratory projects.
A five-year cost checklist
Add up: hood, blower, duct installation and roof penetration; filter replacements over five years; fan energy; annual testing and certification; and the cost of a layout change if the lab's work evolves. The cheapest hood on the quotation is often not the cheapest over the term.