Frame types and when each is used
| C-frame | Open knee space, easy to clean underneath, common for general chemistry and teaching labs. Best where seated work and under-bench access matter. |
|---|---|
| H-frame | Greater lateral stability and higher load capacity. Suited to heavy instruments, long island benches and mobile equipment. |
| Cantilever | Frame hung from a wall or central spine with no front legs. Maximum floor clearance, but requires a verified structural fixing to the building. |
Steel, gauge and finish
Specify the steel type, tube or sheet gauge, weld finish and coating system. For wet areas, stainless steel or a well-sealed powder coat over pre-treated steel is worth the cost. Ask what surface preparation is used before coating — that detail determines how long the finish survives.
Levelling and loads
Adjustable feet are essential on any floor that is not perfectly flat. State the anticipated point load, especially for balance benches, centrifuges and heavy analytical instruments; these should be quoted as separate items with their own frame specification.
Worktops and edges
Phenolic, epoxy resin, ceramic and stainless worktops behave differently under heat, solvents and impact. Specify material, thickness, edge profile (marine edge, raised lip or flat) and joint details. Confirm the chemical resistance data against the chemicals actually used.
Service routing
Show gas, water, power and drainage positions on the shop drawing, plus the service spine or void that carries them. Services fixed after production are the most common cause of site rework.